Mode No.
Function Settings
Class
1 and 2
Class 3
4-421*
1
Dither Pattern
Selection
(Grayscale
Mode/Photo
Mode)
Selects the dither pattern used in
grayscale processing mode.
0: 180-line
1: 140-line
2: 95-line
A greater number of lines gives a more
detailed copy. If the value is changed, the
UP mode setting is also changed.
2 *
Dither Pattern
Selection
(Binary
Picture/Photo
Mode)
Selects the dither pattern used in binary
picture processing mode.
0: 95-line
1: 70-line
2: 140-line
A greater number of lines gives a more
detailed copy. If the value is changed, the
UP mode setting is also changed.
4-902
Exposure
Lamp On
Turns on the exposure lamp.
0: No (Off)
1: Yes (On)
To turn off the exposure lamp, select "0".
4-904*
SBU Gain
Adjustment
Adjusts the coefficient of the D/A
converter for the standard AGC gain curve
0 ~ 255
1/step
30
Do not adjust this value.
However, after
performing the memory all clear
(SP5-801), use it to re-input the previous
value.
4-905*
SBU DC
Count
Adjustment
Adjusts the coefficient of the D/A
converter for the AGC gain curve for dc
count.
0 ~ 255
1/step
30
Do not adjust this value.
However, after
performing the memory all clear
(SP5-801), use it to re-input the previous
value.
4-906*
SBU
Reference
Value
Adjustment
Adjusts the coefficient of the D/A
converter for the AGC gain curve for
scanning the white plate.
0 ~ 255
1/step
147
Do not adjust this value.
However, after
performing the memory all clear
(SP5-801), use it to re-input the previous
value.
4-907*
SBU Offset
Value
Adjustment
Adjusts the coefficient of the D/A
converter for the offset (Z/C) for the
analog image data processing
0 ~ 255
1/step
180
Do not adjust this value.
However, after
performing the memory all clear
(SP5-801), use it to re-input the previous
value.
4-908*
SBU Auto
Adjustment
Performs the auto scanner adjustment.
0: Normal
operation
1: Start the
adjustment
Using this SP mode after replacing the
white plate or the memory on the BICU
board. See "Replacement and Adjustment
- Standard White Level" for details on how
to do this.
4-909*
EDU Test
Mode
Japanese version only.
Do not change the value.
1 August 1996 SERVICE PROGRAM MODE
4-17